Free tool

Salinity & specific-gravity converter

Specific gravity, ppt, a hydrometer read at the wrong temperature — the most muddled number in the hobby. Convert between them, see if you're on the 35 ppt reef target, and correct a glass-hydrometer reading for temperature — the bit most converters quietly skip.

The one thing to remember: salinity (in ppt) is the real target — specific gravity is just a proxy that shifts with temperature and with your instrument's reference. So always know your instrument's reference temperature, and calibrate a refractometer with RO/DI (0) or calibration fluid before you trust it.
Convert salinity ⇄ specific gravity
SG

Assumes the sample was at the instrument's reference temperature. Read a glass hydrometer in warmer tank water? Use the correction below.

Correct a hydrometer read at the wrong temperature
SG

Only works if you KNOW your hydrometer's calibration temperature — a correction with the wrong reference is worse than none (Randy Holmes-Farley). Refractometers with ATC handle this for you.

An honest tool, not a promise. Every value is anchored to Randy Holmes-Farley's published figures (S=35 = 1.0264 at 77 °F; ~1300 ppt per SG unit; seawater density vs temperature), and quoted to about ±0.0002 SG / ±0.5 ppt — finer than any hobby hydrometer. Different salt mixes and ion balances shift this a hair; your calibrated refractometer has the final say.

Sources & method

A free Coralkin tool — no sign-up, nothing saved. It does the conversion; your calibrated instrument has the final say. We never guess a care number.


Salinity ⇄ specific gravity quick reference

Specific gravity shown at the 25 °C / 77 °F reference (most reef refractometers & Tropic Marin). The ~35 ppt reef-ideal band is highlighted.

Salinity (ppt)Specific gravity (SG)
30 1.0226
32 1.0241
33 1.0249 reef-ideal
34 1.0256 reef-ideal
35 1.0264 reef-ideal
36 1.0272

Quick answers

How do I convert specific gravity to ppt (salinity)?

Anchor to natural seawater: a salinity of 35 ppt reads as specific gravity 1.0264 at the 25 °C / 77 °F reference most reef refractometers and Tropic Marin use. From there, every 0.001 change in specific gravity is about 1.3 ppt (Randy Holmes-Farley's cited slope of ~1300 ppt per 1.000 SG). So SG 1.025 is roughly 33.5 ppt at that reference. Enter your reading and this converter does it exactly.

What specific gravity should a reef tank be?

Aim for natural-seawater salinity of about 35 ppt, which is specific gravity ≈1.0264 at 25 °C / 77 °F. Salinity in ppt is the real target; specific gravity is just a temperature-dependent proxy. Corals and inverts are happiest at ~35 ppt — hold it steady rather than chasing an exact SG number.

Why does the same salinity show a different specific gravity?

Because specific gravity is the density of your sample divided by the density of pure water at a reference temperature, so it shifts with which reference your instrument uses. A salinity of 35 reads 1.0264 at 77 °F / 25 °C, 1.0266 at 68 °F / 20 °C, and 1.0269 at 60 °F / 15.6 °C (many lab glass hydrometers). Always know your instrument's reference temperature.

How do I correct a hydrometer reading for temperature?

A glass hydrometer only reads true at its calibration temperature; warmer or cooler water throws it off. The correction multiplies your reading by the ratio of seawater density at the calibration temperature to density at the sample temperature (Randy Holmes-Farley's method) — this tool does it for you. It only works if you KNOW your hydrometer's calibration temperature; a correction with the wrong reference is worse than none. Refractometers with automatic temperature compensation handle this for you.

Should I calibrate my refractometer?

Yes — calibrate a refractometer with RO/DI water (which should read 0) or with calibration fluid before you trust it. Salinity (ppt) is the number that matters, and your calibrated instrument has the final say; different salt mixes and ion balances shift specific gravity by a hair.